Filter sets for checking the photometric linearity

according to Ph. Eur. (EP 10) and USP 857

Article-No. Application Content certified wavelengths in nm
667S501 Checking the photometric linearity in the
UV range
3 filters with Niacin in HCl:
UV506, 6 mg/l (0.25 Abs)
UV512, 12 mg/l (0.5 Abs)
UV524, 24 mg/l (1.0 Abs)
UV599, Hydrochloric acid (HCl), reference filter
213, 261
667350 Checking the photometric linearity in the
UV range
4 filters with Niacin in HCl:
UV506, 6 mg/l (0.25 Abs)
UV512, 12 mg/l (0.5 Abs)
UV518, 18 mg/l (0.75 Abs)
UV524, 24 mg/l (1.0 Abs)
UV599, Hydrochloric acid (HCl), reference filter
213, 261
667307 Checking the photometric linearity in the
UV range
5 filters with Potassium dichromate in HClO4:
UV20, 20 mg/l (0.1 - 0.3 Abs)
UV40, 40 mg/l (0.2 - 0.6 Abs)
UV60, 60 mg/l (0.3 - 0.9 Abs)
UV80, 80 mg/l (0.4 - 1.2 Abs)
UV0100, 100 mg/l (0.5 - 1.45 Abs)
UV14, Perchloric acid (HClO4), reference filter
235, 257, 313, 350
 
666S006 Checking the photometric linearity in the
Vis range
3 Neutral density glass filters:
F2 (0.25 Abs)
F3 (0.5 Abs)
F4 (1.0 Abs)
F0, empty frame
440, 465, 546.1, 590, 635
666S200 Checking the photometric linearity in the
Vis range
3 Neutral density glass filters:
F2 (0.25 Abs)
F4 (1.0 Abs)
F203 (2.0 Abs)
440, 465, 546.1, 590, 635


Request filter set

Reference Materials (Brochure & User Manual EN)

Download


Contact/Inquiry

Hellma USA Inc.
80 Skyline Drive
11803 Plainview, NY
United States

Phone: +1 5169 3908 88
info.us@hellma.com
http://www.hellmausa.com

Contact/Inquiry

Note: Your data will of course be treated confidentially and will not be shared with third parties. You can revoke your consent at any time by clicking on the end of each e-mail.
Required Field