Falcata XP 12

Falcata Xtremely Precise

Ideally-suited for transflection measurements in process applications. The high-performance measuring head enables extremely precise and reliable measurement results. Available in diameters 6 mm and 12 mm.

Suitable for transparent or slightly turbid media

 

  • Reaction monitoring in chemical synthesis
  • Control of process chromatography
  • Determination of color numbers
  • Versatile due to exchangeable path length inserts

 

 

Falcata inserts for path lengths 1 mm, 2 mm, 5 mm, 10 mm and 20 mm

Special Features

Variable path lengths eliminate the need for sample dilution

The Falcata XP model series allows the optical path lengths to be easily adjusted by means of exchangeable path length inserts. This allows different measurements to be performed without the need to dilute the sample.

 

Spectrum of the same sample measured with different path length inserts

Precise and reproducible measurement results

The interchangeable path length tips are screwed to a stop and thus provide a reproducible optical path length.

 

High Process Safety

The probe is composed of standardized components that are kept in stock. This ensures rapid availability of spare parts. This leads to higher process reliability and plant efficiency.

 

Technical Data

Model

Falcata XP 

Measurement Principle

Transflection

Outer Diameter

6 mm I ca. 1/4 inch  or 12 mm | approx.1/2 inch

Optical Path / Focus

1 mm, 2 mm, 5 mm, 10 mm, 20 mm or set (1, 2, 5, 10 and 20 mm)

Optical Tolerance

Quartz ±0.04 mm, Sapphire ±0.04 mm

Probe Barrel Material

Stainless Steel 1.4435/1.4404 (316L) / Hastelloy C-22 (2.4602) / Titan Grade 2 (blazed)

Sealing Technology

Kalrez 4079 / 6230 / 6375 or hard soldered (in combination with fixed path length, sapphire and titan)

Spectral Range

UV/Vis or NIR

Wavelengths

UV/Vis: 230 - 1100 nm; NIR: 400 - 2300 nm

Optical Connection

F-SMA socket and ATEX PMA housing NW23 or 2m glass fiber with PE-coating, 2x F-SMA and PMA-housing NW17

Lightguide Technology

Standard Lighguides / Hightemperature Lightguides

Process Connection

Falcata XP 6 mm: Without flange
Falcata XP 12 mm: Different EN/DIN/ASME flanges / according to customer specification

Probe Barrel

Not Swagelok suitable

Maximum Pressure

40 bar (Class 300, overpressure at RT, 31 bar at 200 °C, 27 bar at 300 °C, 25 bar at 400°C)

Temperature Restrictions

T max: Optical connection 150 °C (without ext. lightguides), Copex PMA 130 °C (85 °C for ATEX)

Maximum Immersion Depth

100 mm - 640 mm (without flange) / 600 mm (with flange)

Minimum Immersion Depth

25 mm

Probe Barrel Length

Immersion depth + 10 mm (without flange), immersion depth + 50 mm (with flange)

Additional Functions

None

Temperature

5 °C at 300 °C (depending on light guide and sealing technology) / blazed at 300 °C

Pressure

-1 bar to 40 bar

 

Optical Probes / Measuring Cells (Brochure EN)

Download

Falcata XP (Datasheet_EN)

Download

Probe and Flow Cell Configurator

Hellma offers the largest selection of products worldwide and the highest level of expertise to support your online measurements from UV to IR. Find the right measuring interface quickly and easily or configure your individual immersion probe or measuring cell.

 

 

You are interested in this product or have any questions?
Please do not hesitate to contact us!

 

E-Mail: sales.analytics@hellma.com
Telephone:  +49 7631 182 10 10

More information about our process probes can be found here.