Falcata Lab

The Falcata Lab probes is ideally suited for numerous laboratory measurements as it has interchangeable path lengths

The small outer diameter (3 mm, 4 mm, 6 mm) allows the probe to be integrated under limited space conditions (e.g laboratory reactors, tablet dissolution tests). The Hellma Analytics micro immersion probe was specifically developed for the measurement of small volumes. Because of its slim design, less sample material is needed than with standard probes. Due to the small outer diameter sample flow is only minimally affected. This probe is very well suited for the measurement of transparent or slightly cloudy media.

Special Features:

  • Flexible use from interchangeable path length tips
  • Windows and mirror are made of the highly purified quartz glass Extended
  • High-transmission optics and marginal scattered light values

Technical Data

Model Falcata Lab
Measuring Principle Transflexion
Outer Diameter 3 /4 /6 mm
Optical Path 1, 2, 5, 10, 20 mm ( Ø 6 mm) or
5, 10 mm (Ø 3 mm, 4 mm)
Opt. Tol. ±0,01 mm
Optical Material Quartz Glass
Sealing Technology Epoxy Adhesion
Probe Body Material Stainless Steel 1.4435 / 1.4404 (316l)  
Spectral Range UV/Vis
NIR
Wavelengths UV/Vis: 210-1100 nm
Device Connection Glass Fiber with F-SMA Connector/ Glass Fiber with Collimator (Ø 5mm)
Fiber Length 2 m glas fiber or no glass fiber
Minimum Immersion Depth 10 mm (Ø 3 mm, 4 mm)
22 mm (Ø 6 mm)
Probe Body Length 75 mm (Ø 3 mm)
130 mm (Ø 4 mm)
Temperature 5 °C to 100 °C
Pressure -1 to 6 bar
Scope of dDelivery Immersion probe, manual, certificate of the pressure test, protocol of the trans-mission test, transport packaging

 